This phantom-based study measured radiation scatter from an ERCP fluoroscopy unit under varying conditions of distance, shielding, and frame rate.
The Permanente Journal · 5 authors, 3 centres
This summary was generated by AI from a single paper. It has not been reviewed by a clinician and is not clinical advice. Verify against the source before acting on it.
This phantom-based study measured radiation scatter from an ERCP fluoroscopy unit under varying conditions of distance, shielding, and frame rate.
Scatter was measured at various distances, with and without a 0.5-mm lead apron, and at frame rates from 2 to 8 fps. An image quality phantom assessed resolution. Increasing distance from 1.5 ft to 9 ft reduced scatter from 0.75 mR/h to 0.15 mR/h (average phantom) and from 50 mR/h to 3.06 mR/h (large phantom). A lead apron reduced scatter from 4.10 to 0.11 mR/h (average phantom) and from 15.30 mR/h to 0.43 mR/h (large phantom). Decreasing frame rate from 8 fps to 2 fps linearly reduced scatter from 55 mR/h to 13.60 mR/h. No change in line pairs was identified across frame rates. Distance measurements deviated from the inverse-square law, likely due to environmental factors. Generalizability may be limited by facility-specific variables including equipment model and room geometry.